Remote control commands
R&S
®
RTE
1201User Manual 1326.1032.02 ─ 20
Parameters:
<AutoTimeScaling> Range: 1E-3 to 1000
Increment: 1
*RST: 0.5
Usage: Asynchronous command
18.9.17 Noise Reject
TRIGger<m>:LEVel<n>:NOISe[:STATe]..........................................................................1201
TRIGger<m>:LEVel<n>:NOISe:MODE........................................................................... 1201
TRIGger<m>:LEVel<n>:NOISe:ABSolute....................................................................... 1202
TRIGger<m>:LEVel<n>:NOISe:PERDivision...................................................................1202
TRIGger<m>:LEVel<n>:NOISe:RELative........................................................................1203
TRIGger<m>:ANEDge:NREJect.................................................................................... 1203
TRIGger<m>:LEVel<n>:NOISe[:STATe] <HysteresisMode>
Selects how the hysteresis is set.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
<n> 1..31
Indicates the trigger source:
1...4 = channel 1 to 4
5 = external trigger input
6...9 and 11 = not available
10 = line trigger
12...27 = R&S RT-ZVC input channels
28...31 = not available
Parameters:
<HysteresisMode> AUTO | MANual
AUTO
This is the recommended mode. The hysteresis is set by the
instrument to reject at least the internal noise of the instrument.
You can define a higher minimum value using TRIGger<m>:
LEVel<n>:NOISe:ABSolute.
MANual
The hysteresis is defined directly with TRIGger<m>:
LEVel<n>:NOISe:ABSolute.
*RST: AUTO
Usage: Asynchronous command
TRIGger<m>:LEVel<n>:NOISe:MODE <HystMode>
Selects how the hysteresis is set.
Trigger