Remote control commands
R&S
®
RTE
1794User Manual 1326.1032.02 ─ 20
BUS<m>:USB:DIFFerential:THReshold <ThresDiff>
Defines the threshold level for the differential signal in the USB high speed protocol.
Suffix:
<m>
.
1..4
Parameters:
<ThresDiff> Range: -2 to 2
Increment: 0.01
*RST: 0
Default unit: V
BUS<m>:USB:DATA:THReshold
<ThresholdData>
Defines the threshold level for the data signal in the USB HSIC protocol.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdData> Range: -2 to 2
Increment: 0.01
*RST: 0.65
Default unit: V
BUS<m>:USB:STRobe:THReshold <ThresholdStrobe>
Defines the threshold level for the strobe signal in the USB HSIC protocol.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdStrobe> Range: -2 to 2
Increment: 0.01
*RST: 0.65
Default unit: V
18.17.17.2 Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
TRIGger<m>:SOURce[:SELect] is set to SBUS.
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
Decoding is enabled: BUS<m>[:STATe] is set to ON.
Protocols

Table of Contents

Other manuals for Rohde & Schwarz RTE

Related product manuals