Acquisition and waveform setup
R&S
®
RTE
218User Manual 1326.1032.02 ─ 20
Source Channels
You can simultaneously connect a R&S RT-ZVC and a R&S RT-ZL04 to the R&S RTE,
but no parallel operation on screen is possible.
You can acquire and measure the R&S RT-ZVC or R&S RT-ZL04 together with the
analog input channels. They are running on the same horizontal scale.
If an amperemeter or voltmeter channel is activated, it can be displayed on the screen
and used as a source for:
Cursor measurements
Automatic measurements, and also histogram, limit checks, longterm, track, quick
measurements
Mask testing
Basic and advanced maths
Reference waveform
Trigger type "Edge"
Search: all search events
XY diagrams
Data export
All features of the base unit and the R&S RTE-K18 option are supported.
Data export
You can save the data of the amperemeter and voltmeter channels to an XML, CSV, or
BIN file. One channel per file can be saved. Files in BIN format can be reloaded to the
R&S RTE as reference waveforms.
See also:
Chapter 12.2.7, "Saving and loading waveform data", on page 496
Chapter 12.2.2, "Waveforms - export settings", on page 485
Remote commands for export to file:
EXPort:WAVeform:SOURce on page 1400
EXPort:WAVeform:NAME on page 1402
EXPort:WAVeform:SAVE on page 1403
5.6.1 R&S RT-ZVC overview
5.6.1.1 Setup
Access: "Vertical" menu > "ZVC Multi-Channel Probe" > "Setup" tab
R&S
RT-ZVC probe