Remote control commands
R&S
®
RTE
1867User Manual 1326.1032.02 ─ 20
Return values:
<FrameByteValue> Range: 0 to 65535
Increment: 1
*RST: 0
Usage: Query only
18.17.19.4 Search settings
SEARch:TRIGger:SWIRe:CTYPe.................................................................................. 1867
SEARch:TRIGger:SWIRe:DATA:CONDition.................................................................... 1867
SEARch:TRIGger:SWIRe:DATA:MAX.............................................................................1868
SEARch:TRIGger:SWIRe:DATA:MIN..............................................................................1868
SEARch:TRIGger:SWIRe:ERRor:ESC........................................................................... 1868
SEARch:TRIGger:SWIRe:ERRor:PARity........................................................................ 1869
SEARch:TRIGger:SWIRe:TIME:CONDition.....................................................................1869
SEARch:TRIGger:SWIRe:TIME:MAX.............................................................................1869
SEARch:TRIGger:SWIRe:TIME:MIN.............................................................................. 1870
SEARch:TRIGger:SWIRe:TYPE.................................................................................... 1870
SEARch:TRIGger:SWIRe:CTYPe <SearchName>,<ControlType>
SEARch:TRIGger:SWIRe:CTYPe? <SearchName>
Searches for a specific control type character.
Parameters:
<ControlType> ANY | FCT | EOP | EEP
FCT
Flow Control Token
EOP
Normal End of Packet
EEP
Error End of Packet
*RST: ANY
Parameters for setting and query:
<SearchName>
SEARch:TRIGger:SWIRe:DATA:CONDition <SearchName>,<DataOperator>
SEARch:TRIGger:SWIRe:DATA:CONDition? <SearchName>
Set the condition for the data pattern data value. You can define an exact value or a
value range.
Parameters:
<DataOperator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
Protocols