Protocol analysis
R&S
®
RTE
734User Manual 1326.1032.02 ─ 20
The graphs below illustrate the variants.
Table 13-14: Enhanced serial message formats
16 bit data and 4 bit message ID 12 bit data and 8 bit message ID
Trigger
The R&S RTE can trigger on various parts of SENT pulses. The data line must be con-
nected to an input channel, triggering on math and reference waveforms is not possi-
ble.
SENT enables you to trigger on:
Calibration/synchronization pulse
Transmission sequence
Serial messages
Error conditions
13.13.2 SENT configuration
13.13.2.1 SENT configuration
Access: [PROTOCOL] > "Decode" tab > "Protocol" = SENT
Make sure that the tab of the correct serial bus is selected.