457User Manual 1326.1032.02 ─ 20
Clock source, Clock edge, Clock level
Set the clock settings. Both "Clock level" and "Clock edge" define the starting point for
calculation of the setup and hold time.
Remote command:
SEARch:TRIGger:DATatoclock:CSOurce on page 1370
SEARch:TRIGger:DATatoclock:CEDGe on page 1369
SEARch:TRIGger:DATatoclock:CLEVel on page 1370
Data level
Sets the voltage level for the data signal. At this level, the setup and hold time are
measured.
Remote command:
SEARch:TRIGger:LEVel[:VALue] on page 1355
Trigger level
Sets the voltage level for trigger search. The value is used for all search event types
that require one trigger level. The search trigger level is search-specific, you can define
different levels in different searches for the same event.
Remote command:
SEARch:TRIGger:LEVel[:VALue] on page 1355
Setup time, Hold time
Sets the minimum time before (Setup) and after (Hold) the clock edge while the data
signal must stay steady above or below the data level.
See also: "Setup time" on page 255 and "Hold time" on page 255.
Remote command:
SEARch:TRIGger:DATatoclock:STIMe on page 1371
SEARch:TRIGger:DATatoclock:HTIMe on page 1370
11.2.2.10 State
The state search is a qualified edge search. The state search is only available for ana-
log channel sources (Ch1 to Ch4).
Search setup