Remote control commands
R&S
®
RTE
1598User Manual 1326.1032.02 ─ 20
18.17.9.2 Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
TRIGger<m>:SOURce[:SELect] is set to SBUS.
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:I2S:TYPE................................................................................................ 1598
TRIGger<m>:I2S:TCONdition<n>:CHANnel....................................................................1599
TRIGger<m>:I2S:TCONdition<n>:CONDtion...................................................................1599
TRIGger<m>:I2S:TCONdition<n>:DMIN.........................................................................1600
TRIGger<m>:I2S:TCONdition<n>:DMAX........................................................................1600
TRIGger<m>:I2S:SOWords...........................................................................................1601
TRIGger<m>:I2S:WSSLope.......................................................................................... 1601
TRIGger<m>:I2S:TYPE <Type>
Selects the trigger type for audio signal analysis.
Parameters:
<Type> DATA | WINDow | CONDition | WSELect | ECONdition
DATA
Triggers on a data word or data range on a specified channel or
on any channel.
To set the channel, use TRIGger<m>:I2S:TCONdition<n>:
CHANnel.
To set the data condition, use:
TRIGger<m>:I2S:TCONdition<n>:CONDtion,
TRIGger<m>:I2S:TCONdition<n>:DMIN and TRIGger<m>:
I2S:TCONdition<n>:DMAX.
WINDow
Triggers if the decoded data values stay inside a "window" that
is formed by a data range and a time specified by a number of
subsequent words. It considers a selected channel or all chan-
nels.
To set up a window trigger, you define the channel and data con-
dition in the same way as for DATA trigger type. Additionally, you
set the time limit with TRIGger<m>:I2S:SOWords.
CONDition
The frame condition trigger sets the trigger on an AND combina-
tion of data conditions on different channels. The instrument trig-
gers if all conditions are met inside one frame.
To set up a CONDition trigger, you define up to four channel and
data conditions in the same way as for DATA trigger type.