Remote control commands
R&S
®
RTE
1155User Manual 1326.1032.02 ─ 20
Parameters:
<UseZeroComp> ON | OFF
*RST: OFF
Usage: Asynchronous command
ZVC:Z<m>:I<n>:ZERComp:STATe?
Returns the status of the zero compensation of the indicated current channel.
Suffix:
<m>
.
1..2
<n> 1..4
Return values:
<ZeroCompState> ON | OFF
ON: Zero compensation is used
Off: Zero Compensation is not used
*RST: OFF
Usage: Query only
Asynchronous command
18.8.8.2 Resolution and data
ACQuire:POINts:ZVALue?.............................................................................................1155
ACQuire:ZRESolution?................................................................................................. 1155
ZVC:Z<m>:I<n>:DATA:HEADer?....................................................................................1156
ZVC:Z<m>:V<n>:DATA:HEADer?.................................................................................. 1156
ZVC:Z<m>:I<n>:DATA[:VALues]?...................................................................................1156
ZVC:Z<m>:V<n>:DATA[:VALues]?................................................................................. 1156
ACQuire:POINts:ZVALue?
Returns the current record length used by the R&S RT-ZVCmulti-channel power probe
channel.
Return values:
<ZUIRecordLength> Range: 1000 to 400E6
Increment: 2
*RST: 1000
Default unit: pts
Usage: Query only
Asynchronous command
ACQuire:ZRESolution?
Returns the current resolution of the R&S RT-ZVCmulti-channel power probe channel.