EasyManua.ls Logo

Rohde & Schwarz RTE - Trigger

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1651User Manual 1326.1032.02 ─ 20
T100
Sets the thresholds to the default values for 10BASE-T
(100 meters): upper threshold to 0.75 V, lower threshold to
-0.75 V
TX0
Sets the thresholds to the default values for 100BASE-TX
(0 meters): upper threshold to 0.5 V, lower threshold to -0.5 V
TX100
Sets the thresholds to the default values for 100BASE-TX
(100 meters): upper threshold to 0.35 V, lower threshold to
-0.35 V
MANual
Allows to set individual threshold voltage levels
*RST:
T0
BUS<m>:ETHernet:BITRate
<BitRateValue>
Sets the bit rate value that defines the transmission speed in bits per second.
Suffix:
<m>
.
1..4
Parameters:
<BitRateValue> Range: 10000 to 150000000
Increment: 1000
*RST: 10000000
Default unit: bps
18.17.12.2 Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
TRIGger<m>:SOURce[:SELect] is set to SBUS.
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
Decoding is enabled: BUS<m>[:STATe] is set to ON.
TRIGger<m>:ETHernet:TYPE....................................................................................... 1652
TRIGger<m>:ETHernet:PATTern....................................................................................1652
TRIGger<m>:ETHernet:ERRor:CRC..............................................................................1652
TRIGger<m>:ETHernet:ERRor:LENGth..........................................................................1653
TRIGger<m>:ETHernet:ERRor:PREamble......................................................................1653
TRIGger<m>:ETHernet:FRAMe:CCONdition...................................................................1653
TRIGger<m>:ETHernet:FRAMe:CMAX.......................................................................... 1654
TRIGger<m>:ETHernet:FRAMe:CMIN............................................................................1654
TRIGger<m>:ETHernet:FRAMe:DCONdition...................................................................1654
TRIGger<m>:ETHernet:FRAMe:DMAX.......................................................................... 1654
Protocols