Measurements
R&S
®
RTE
355User Manual 1326.1032.02 ─ 20
Table 8-4: Counting measurement types
Meas. type Symbol Description/Result
Pulse count
The number of positive or negative pulses of the waveform, or of
both positive and negative pulses.
The mean value of the signal is determined. If the signal passes
the mean value, an edge is counted. A positive pulse is counted
if a rising edge and a following falling edge are detected. A neg-
ative pulse is counted if a falling edge and a following rising
edge are detected.
Edge count
The number of positive or negative edges, or of both positive
and negative edges.
The instrument determines the mean value of the signal and
counts an edge every time the signal passes the mean value.
8.2.5.2 Settings for amplitude/time measurements
Access: "Meas" menu > "Setup" > "Amp/Time" category
Amplitude and time measurements are available for sources in the time domain. For
some amplitude/time measurements, such as delay, setup/hold and delay to trigger,
further setting are required to get a measurement result.
This chapter explains all settings for amplitude/time measurements. For a description
of available measurement type, see Chapter 8.2.5.1, "Overview of amplitude/time mea-
surements", on page 349.
Envelope..................................................................................................................... 355
Pulses slope................................................................................................................356
Measured slope...........................................................................................................356
Settings for delay and phase measurements (analog sources).................................. 356
Settings for delay and phase measurements (digital sources)................................... 357
Setup/Hold measurement settings..............................................................................357
Clock ref level..............................................................................................................358
Data ref level...............................................................................................................358
Pulse count................................................................................................................. 359
Edges slope................................................................................................................ 359
Delay to trigger measurement settings....................................................................... 359
Envelope
This setting is only available for measurements on envelope waveforms, see "Arith-
metic" on page 182.
Automatic measurements