Measurements
R&S
®
RTE
356User Manual 1326.1032.02 ─ 20
"Both"
The upper and the lower envelope are used in measurements. For
time measurements, the averages of min and max values are used,
that is, the measurement is performed on the average waveform built
from the upper and lower envelope.
"Maximum"
Measurements are performed on the upper envelope.
"Minimum"
Measurements are performed on the lower envelope.
Remote command:
MEASurement<m>:ENVSelect on page 1259
Pulses slope
Sets the first slope of the pulses to be counted.
The setting is available only for the "Pulse count" measurement.
"Positive"
Positive pulses are counted.
"Negative"
Negative pulses are counted.
"Either"
Both positive and negative pulses are counted.
Remote command:
MEASurement<m>:AMPTime:PSLope on page 1261
Measured slope
Selects the slope direction for frequency and period measurements.
"Positive /
Negative"
Measures the time between rising or falling edges, respectively.
"Either"
In multiple measurements, the time is measured both between rising
edges and between falling edges.
In single measurements. The first edge is taken for the measurement.
"First edge"
Time is measured either between rising edges or between falling
edges. The first edge is taken for the measurement. In single mea-
surements, it works the same way as "Either".
Only available for analog measurement sources.
Remote command:
MEASurement<m>:AMPTime:PFSLope on page 1260
Settings for delay and phase measurements (analog sources)
You can measure the time between any two slopes at middle reference level. There-
fore, the slopes must be defined for each source individually. The measurement result
is negative if the edge of the second source comes before the edge of the first source.
With the settings shown in the picture, the time between the second and the fifth rising
edge is measured.
Automatic measurements