Remote control commands
R&S
®
RTE
1261User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<PeriodSlope> FIRSt | POSitive | NEGative | EITHer
POSitive | NEGative
Measures the time between rising or falling edges, respectively.
EITHer
In multiple measurements, the time is measured both between
rising edges and between falling edges.
In single measurements. the first edge is taken for the measure-
ment.
FIRSt
Time is measured either between rising edges or between falling
edges. The first edge is taken for the measurement. In single
measurements, it works the same way as "Either".
Only available for analog measurement sources.
*RST:
FIRSt (analog source), POSitive (digital source)
Usage: Asynchronous command
MEASurement<m>:AMPTime:PSLope <PulsesSlope>
Sets the first slope of the pulses to be counted. The setting is only relevant for pulse
count measurement (MEASurement<m>:MAIN PULCnt or
MEASurement<m>:ADDitional PULCnt,ON).
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Parameters:
<PulsesSlope> POSitive | NEGative | EITHer
Count either positive or negative pulses, or both.
*RST: POSitive
Usage: Asynchronous command
MEASurement<m>:AMPTime:DELay<n>:DIRection <EdgeCntDirct>
Selects the direction for counting slopes for each source: from the beginning of the
waveform, or from the end.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Automatic measurements