Protocol analysis
R&S
®
RTE
519User Manual 1326.1032.02 ─ 20
Standard lines
Standard lines define the contents of the label list. The rules for standard lines follow
the csv convention, they are:
Values are separated by commas
Space characters following a delimiter are ignored
Values with a special character (comma, newline, or double quote) must be
enclosed in double quotes
Text in double quotes must be escaped by double quote characters
The format of the numeric value is indicated by a suffix. The following formats are sup-
ported:
Format Suffix Example
Decimal <empty>
d
106, DeviceName
106d, DeviceName
Hexadecimal h
6Ah, DeviceName
or prefix: 0x6A, DeviceName
Octal o
152o, DeviceName
Binary b
01101010b, DeviceName
The maximum supported word size for (unsigned) integers is 64 bits.
# --- Start of PTT file
@FILE_VERSION = 1.0
@PROTOCOL_NAME = i2c
# Following two lines are equal:
7,01h,Temperature
7,01h, Temperature
# A comma must be enclosed in double quotes:
7,01h,"Temperature, Pressure, and Volume"
# A double quote must also be enclosed in double quotes:
7,7Fh,"Highspeed ""Master"" 01"
# Following lines yield the same result:
7d,0x11,Pressure
7h,11h,Pressure
0x7,17d,Pressure
7,17,Pressure
13.1.4.2 Label list - general settings
In the "Label List" tab, you can load, read and activate label list files.
Basics of protocol analysis