Remote control commands
R&S
®
RTE
1359User Manual 1326.1032.02 ─ 20
Usage: Asynchronous command
SEARch:TRIGger:INTerval:WIDTh <SearchName>,<Width>
SEARch:TRIGger:INTerval:WIDTh? <SearchName>
Defines the time between two pulses.
Parameters:
<Width> Range: 100E-12 to 864
Increment: 100E-9
*RST: 5E-9
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
18.15.6 Runt search conditions
SEARch:TRIGger:RUNT:DELTa.....................................................................................1359
SEARch:TRIGger:RUNT:POLarity..................................................................................1359
SEARch:TRIGger:RUNT:RANGe................................................................................... 1360
SEARch:TRIGger:RUNT:WIDTh.................................................................................... 1360
SEARch:TRIGger:LEVel:RUNT:LOWer...........................................................................1361
SEARch:TRIGger:LEVel:RUNT:UPPer........................................................................... 1361
SEARch:TRIGger:RUNT:DELTa <SearchName>,<WidthDelta>
SEARch:TRIGger:RUNT:DELTa? <SearchName>
Defines a range around the given runt width.
Parameters:
<WidthDelta> Range: 100E-12 to 864
Increment: 100E-9
*RST: 100E-12
Default unit: s
Parameters for setting and query:
<SearchName> Search definition
Usage: Asynchronous command
SEARch:TRIGger:RUNT:POLarity <SearchName>,<Polarity>
SEARch:TRIGger:RUNT:POLarity? <SearchName>
Indicates the polarity of a pulse, that is the direction of the first pulse slope.