Mixed signal option (MSO, R&S RTE-B1)
R&S
®
RTE
914User Manual 1326.1032.02 ─ 20
Remote command:
TRIGger<m>:PARallel:TIMeout:RANGe on page 1915
Time
Defines the time limit for the timeout at which the instrument triggers.
Remote command:
TRIGger<m>:PARallel:TIMeout:TIME on page 1916
Logical expression
Defines a logic combination of several digital channels as trigger condition if "Logic" is
set for "Source". The "Qualification Editor" supports the entry of the expression.
Remote command:
TRIGger<m>:PARallel:TIMeout:EXPRession[:DEFine] on page 1913
TRIGger<m>:PARallel:STATe:EXPRession[:DEFine] on page 1913
TRIGger<m>:PARallel:PATTern:EXPRession[:DEFine] on page 1913
TRIGger<m>:PARallel:SPATtern:EXPRession[:DEFine] on page 1913
14.3.1.5 Data2Clock
The "Data2Clock" trigger event occurs when the state of the trigger source signal
changes inside a given time before the clock edge (setup time) or after the clock edge
(hold time). The trigger event occurs at the clock edge for which the setup and/or hold
time was violated.
This trigger type is also known as setup/hold trigger.
With "Data2Clock" trigger, you can trigger on a single digital channel, or a parallel bus
to check several digital channels simultaneously. The clock signal is connected to one
of the digital channels.
If you configure this trigger type for a parallel bus, the bus configuration is adjusted by
the instrument if necessary. The bus is defined as clocked bus, and the clock source of
the trigger is set as clock source of the bus.
Figure 14-4: Data2Clock trigger settings
Clock source
Selects the digital channel of the clock signal.
Trigger