Remote control commands
R&S
®
RTE
1492User Manual 1326.1032.02 ─ 20
BUS<m>:CAN:FDATa:T2Segment <FlexDatTimeSeg2>
BUS<m>:CAN:T2Segment <TimeSeg2>
Sets the number of time quanta after the sample point (T2Segment). T2Segment
matches Phase_seg2 specified in the CAN standard.
Make sure to set also BUS<m>:CAN:T1Segment on page 1491 for correct definition of
the sample point. Alternatively, you can use BUS<m>:CAN:SAMPlepoint.
See also: "Synchronization: Sample point, Time segments, Jump width" on page 565
For CAN FD signals, BUS<m>:CAN:T2Segment defines the synchronization of the
arbitration phase, and BUS<m>:CAN:FDATa:T2Segment defines the synchronization
of the data phase.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<FlexDatTimeSeg2>
<TimeSeg2>
Time quanta
Range: 1 to 21
Increment: 1
*RST: 3.4
BUS<m>:CAN:FDATa:JWIDth <FlexDatJumpWdt>
BUS<m>:CAN:JWIDth <JumpWidth>
Defines the maximum number of time quanta for phase correction. Time segment1
may be lengthened or Time segment2 may be shortened due to resynchronization.
Resynchronization corrects the phase error of an edge caused by the drift of the oscil-
lators.
For CAN FD signals, this setting defines the synchronization of the arbitration phase.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<JumpWidth> Time quanta
Range: 1 to 4, available maximum depends on the num-
ber of segments and the sample point
Increment: 1
*RST: 1
18.17.6.2 Trigger
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
TRIGger<m>:SOURce[:SELect] is set to SBUS.
Protocols