Acquisition and waveform setup
R&S
®
RTE
186User Manual 1326.1032.02 ─ 20
Make sure that the correct channel tab is selected. The vertical rotary knobs are illumi-
nated in the color of the selected channel.
Show channel..............................................................................................................186
Ground........................................................................................................................ 186
Vertical scale...............................................................................................................186
Position....................................................................................................................... 186
Invert channel..............................................................................................................187
Coupling......................................................................................................................187
Offset...........................................................................................................................187
Bandwith..................................................................................................................... 188
Show channel
Switches the channel signal on or off. The signal icon appears on the signal bar. The
waveform of the last acquisition is displayed in the diagram.
Remote command:
CHANnel<m>:STATe on page 1117
Ground
Connects the input to the ground.
Remote command:
CHANnel<m>:GND on page 1117
Vertical scale
Defines the vertical scale in Volts per division. Increasing the scale compresses the
display of the signal.
Remote command:
CHANnel<m>:SCALe on page 1118
Position
Moves the selected signal up or down in the diagram. The visual effect is the same as
for Offset but the waveform is adjusted later in the signal flow. While the offset sets a
voltage, position is a graphical setting given in divisions.
Vertical settings