Mask testing
R&S
®
RTE
442User Manual 1326.1032.02 ─ 20
10.3.4 Running a mask test
Before you can start a mask test, make sure that the mask setup is complete:
●
The mask is defined, see Chapter 10.3.1.1, "Creating user masks", on page 436
and Chapter 10.3.1.2, "Modifying user masks", on page 437.
●
The mask test is defined, see Chapter 10.3.2, "Setting up a mask test",
on page 440
●
The mask display is configured, see Chapter 10.3.3, "Configuring the mask and hit
display", on page 441.
You can perform continuous testing or test a specified number of acquisitions.
1. Press the [MASKS] key on the front panel.
2. Select the "Test Definition" tab.
3. Select "Enable test".
If the acquisition is running, the test starts immediately.
4. If the acquisition is not running, press [RUN CONT].
The test starts and runs until you stop the acquisition or the stop action is executed
if defined.
5. To test a specified number of acquisitions:
a) Press the [ACQUISITION] key.
b) Set the "Average count" to the number of acquisitions.
See also:"Average count (N-single count)" on page 182
c) Press [RUN N× SINGLE].
Working with masks