Remote control commands
R&S
®
RTE
1123User Manual 1326.1032.02 ─ 20
[<Length>] Number of waveform points to be retrieved.
Range: 1 to n. Limit: n + m >= record length
Return values:
<Data> List of values according to the format and content settings.
Example: Retrieve the complete channel 1 waveform, only Y-values:
FORM ASC
EXP:WAV:INCX OFF
CHAN1:WAV1:DATA?
<-- -0.125000,-0.123016,-0.123016,-0.123016,-0.123016,-0.123016,...
Example: Retrieve the first 10 values of the waveform:
CHANnel:WAVeform:DATA:VALues? 0,10
<-- -0.10079051554203,-0.098814234137535,-0.098814234137535,
-0.096837945282459,-0.094861663877964,-0.094861663877964,
-0.092885382473469,-0.090909093618393,-0.090909093618393,
-0.088932812213898
Example: Skip 5 samples and retrieve the next 5 samples:
CHANnel:WAVeform:DATA:VALues? 5,5
<-- -0.094861663877964,-0.092885382473469,-0.090909093618393,
-0.090909093618393,-0.088932812213898
Usage: Query only
18.8.7 Probes
TRPRobe:... command are dedicated commands for the external trigger input.
● Common probe settings...................................................................................... 1123
● Micro button and R&S ProbeMeter..................................................................... 1126
● Passive probes....................................................................................................1131
● Active voltage probes..........................................................................................1132
● Modular probes................................................................................................... 1135
● Predefined probes...............................................................................................1138
● Current probes.................................................................................................... 1139
● Probe attributes...................................................................................................1140
● Probe adapter R&S RT‑Z2T................................................................................1142
18.8.7.1 Common probe settings
TRPRobe:SETup:STATe?..............................................................................................1124
PROBe<m>:SETup:STATe?.......................................................................................... 1124
TRPRobe:SETup:TYPE?.............................................................................................. 1124
PROBe<m>:SETup:TYPE?........................................................................................... 1124
TRPRobe:SETup:NAME?..............................................................................................1124
PROBe<m>:SETup:NAME?.......................................................................................... 1124
TRPRobe:SETup:BANDwidth?...................................................................................... 1125
PROBe<m>:SETup:BANDwidth?................................................................................... 1125
TRPRobe:SETup:ATTenuation[:AUTO]?......................................................................... 1125
Acquisition and setup