Remote control commands
R&S
®
RTE
1485User Manual 1326.1032.02 ─ 20
TRIGger<m>:UART:FCONdition <DataOperator>
Selects the operator for the data pattern (TRIGger<m>:UART:DATA).
Parameters:
<DataOperator> EQUal | NEQual
*RST: EQUal
TRIGger<m>:UART:DATA <Data>
Specifies the data pattern to be found on the specified trigger source, in binary or hex
format. Enter the words in msb first bit order.
Parameters:
<Data> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
Usage: Asynchronous command
18.17.5.3 Decode results
BUS<m>:UART:WORD<n>:RXValue?............................................................................1485
BUS<m>:UART:WORD<n>:TXValue?............................................................................ 1485
BUS<m>:UART:WORD<n>:BITRate?............................................................................ 1486
BUS<m>:UART:WORD<n>:COUNt?..............................................................................1486
BUS<m>:UART:WORD<n>:SOURce?............................................................................1486
BUS<m>:UART:WORD<n>:STARt?............................................................................... 1487
BUS<m>:UART:WORD<n>:STATe?............................................................................... 1487
BUS<m>:UART:WORD<n>:RXValue?
BUS<m>:UART:WORD<n>:TXValue?
Returns the value of the specified word on the Rx line or Tx line, respectively.
Suffix:
<m>
.
1..4
Selects the serial bus.
<n> *
Selects the word.
Return values:
<Value> To set the value format, use FORMat:BPATtern.
The stated values for range, increment and reset are decimal
values.
Range: 0 to 511
*RST: 0
Usage: Query only
Protocols