Remote control commands
R&S
®
RTE
1206User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<State> ON | OFF
*RST: OFF
Usage: Asynchronous command
TRIGger<m>:SEQuence:RESet:TIMeout:TIME
<ResetTimeout>
The time the instrument waits for the number of B-events specified using
TRIGger<m>:SEQuence:COUNt before the sequence is restarted with the A-event.
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<ResetTimeout> Range: 0 to 50
Increment: 1E-12
*RST: 0
Default unit: s
Usage: Asynchronous command
18.9.19 Trigger control
TRIGger<m>:MODE.....................................................................................................1206
TRIGger<m>:FORCe................................................................................................... 1207
TRIGger<m>:OUT:STATe..............................................................................................1207
TRIGger<m>:OUT:POLarity.......................................................................................... 1207
TRIGger<m>:OUT:PLENgth.......................................................................................... 1208
TRIGger<m>:OUT:DELay............................................................................................. 1208
TRIGger<m>:MODE <TriggerMode>
Sets the trigger mode which determines the behaviour of the instrument if no trigger
occurs.
See also: "Trigger mode" on page 268
Suffix:
<m>
.
1..3
Irrelevant, omit the suffix.
Parameters:
<TriggerMode> AUTO | NORMal | FREerun
Trigger