Remote control commands
R&S
®
RTE
1135User Manual 1326.1032.02 ─ 20
Parameters:
<Sound> ON | OFF
*RST: OFF
Usage: Asynchronous command
18.8.7.5 Modular probes
The commands of this chapter are relevant for R&S RT-ZM modular probes. The com-
mands are available in firmware version 3.40 and higher.
The suffix <m> selects the input channel to which the probe is connected.
TRPRobe:SETup:PRMode............................................................................................ 1135
PROBe<m>:SETup:PRMode.........................................................................................1135
PROBe<m>:SETup:DMOFfset.......................................................................................1136
PROBe<m>:SETup:CMOFfset.......................................................................................1136
PROBe<m>:SETup:NOFFset........................................................................................ 1136
PROBe<m>:SETup:POFFset.........................................................................................1137
PROBe<m>:SETup:TERM:STATe.................................................................................. 1137
PROBe<m>:SETup:TERM:MODE..................................................................................1137
PROBe<m>:SETup:TERM:MEASure?............................................................................1138
PROBe<m>:SETup:TERM:ADJust.................................................................................1138
TRPRobe:SETup:PRMode <PrbMeasMd>
PROBe<m>:SETup:PRMode <MeasMode>
Sets the measurement mode of the modular probe.
Suffix:
<m>
.
1..4
Parameters:
<MeasMode> DMODe | CMODe | PMODe | NMODe
DMODe
Differential mode input voltage (V
dm
), the voltage between the
positive and negative input terminal.
CMODe
Common mode input voltage (V
cm
) , the mean voltage between
the positive and negative input terminal vs. ground.
PMODe
Positive single-ended input voltage (V
p
). the voltage between the
positive input terminal and ground.
NMODe
Negative single-ended input voltage (V
N
). the voltage between
the negative input terminal and ground.
*RST:
DMODe
Usage: Asynchronous command