Protocol analysis
R&S
®
RTE
787User Manual 1326.1032.02 ─ 20
13.14.2.4 Custom: Manchester / NRZ display settings
Access: [PROTOCOL] > "Display" tab.
For common display settings, see Chapter 13.1.3, "Display", on page 515.
To enhance the decode possibilities of the custom serial protocol, you can use an addi-
tional setting in the "Display" tab: "Decode layer".
Decode layer
Selects the decode layer.
Decoding is performed in several steps, and the end results are presented in the
decode table. The decode layer selects an interim step for which the decoding result is
shown in the honeycomb display.
"Final"
...
"Edges"
...
"Binary"
...
"Synchroniza-
tion"
...
13.14.2.5 Configuring custom Manchester / NRZ signals
For configuration, assign the lines to the input channels, define the active states and
the logical thresholds, and specify frame format descriptions.
Serial bus setup
1. Press the [PROTOCOL] key on the front panel.
2. Select the tab of the bus you want to set up, for example "SB1".
3. In the "Decode" tab, select the protocol: "Custom".
4. Switch to the "Trigger" tab, tap "Source" and select "Serial bus".
This prevents using digital waveforms (Math and Reference) as channel signals.
Custom: Manchester / NRZ (option R&S
RTE-K50)M