Mixed signal option (MSO, R&S RTE-B1)
R&S
®
RTE
913User Manual 1326.1032.02 ─ 20
Width
For the ranges "Shorter" and "Longer", the width defines the maximum and minimum
pulse width, respectively.
For the ranges "Within" and "Outside", the width defines the center of a range which is
defined by the limits "±Delta".
Remote command:
TRIGger<m>:PARallel:WIDTh:WIDTh on page 1915
±Delta
Defines a range around the given width value.
The combination "Range" = Within and "±Delta" = 0 triggers on pulses with a pulse
width that equals "Width".
The combination "Range" = Outside and "±Delta" = 0 means to trigger on pulse widths
≠ "Width".
Remote command:
TRIGger<m>:PARallel:WIDTh:DELTa on page 1915
14.3.1.4 Timeout
The timeout trigger event checks if the trigger source signal stays above or below the
threshold voltage for a specified time lapse. In other words, the event occurs if the
state condition remains unchanged for the specified time.
You can use the timeout trigger on a single digital channel, or a logical combination of
digital channels.
Figure 14-3: Timeout trigger settings for trigger source = logical combination of digital channels
Range
Sets the state condition:
"Stays high"
The level of a digital channel stays above the threshold, or the logical
expression for "Logic" trigger source is true.
"Stays low"
The level of a digital channel stays below the threshold, or the logical
expression for "Logic" trigger source is false.
"High or low"
The signal state remains unchanged.
Trigger