Acquisition and waveform setup
R&S
®
RTE
204User Manual 1326.1032.02 ─ 20
5.5.4.3 R&S RT-ZPR power rail probes
Figure 5-11: Probe setup for power rail probe R&S
RT-ZPR
The offset is the channel offset. See "Offset" on page 187.
If the ProbeMeter is active, an additional function is provided in the "ProbeMeter" result
box:
Figure 5-12: ProbeMeter result box for
Specific settings for R&S RT-ZPR probes are the following:
AC Coupling
Enables AC coupling in the R&S RT-ZPR power rail probes, which removes DC and
very low-frequency components. The R&S RT-ZPR probe requires 50 Ω input termina-
tion, for which the channel AC coupling is not available. The probe setting allows AC
coupling also at 50 Ω inputs.
Remote command:
PROBe<m>:SETup:ACCoupling on page 1133
TRPRobe:SETup:ACCoupling on page 1133
Probes