Remote control commands
R&S
®
RTE
1185User Manual 1326.1032.02 ─ 20
The hold time can be negative. In this case, the setup time is always positive. The
setup/hold interval starts before the clock edge (setup time) and ends before the clock
edge (hold time). If you change the negative hold time, the setup time is adjusted by
the intrument.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<HoldTime> Range: -99.999E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
TRIGger<m>:DATatoclock:STIMe
<SetupTime>
Sets the minimum time before the clock edge while the data signal must stay steady
above or below the data level.
The setup time can be negative. In this case, the hold time is always positive. The
setup/hold interval starts after the clock edge (setup time) and ends after the clock
edge (hold time). If you change the negative setup time, the hold time is adjusted by
the instrument.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<SetupTime> Range: -99.999E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
18.9.11 State trigger
The state trigger combines the edge trigger settings with trigger qualification. It is only
available for the A-event (Suffix = 1).
Use the following commands:
●
TRIGger<m>:EDGE:SLOPe on page 1168
●
TRIGger<m>:LEVel<n>[:VALue] on page 1166
●
TRIGger<m>:SCOupling on page 1184
TRIGger<m>:QUALify<n>:A[:ENABle] <State>
TRIGger<m>:QUALify<n>:B[:ENABle] <State>
Trigger