EasyManua.ls Logo

Rohde & Schwarz RTE - State Trigger

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1185User Manual 1326.1032.02 ─ 20
The hold time can be negative. In this case, the setup time is always positive. The
setup/hold interval starts before the clock edge (setup time) and ends before the clock
edge (hold time). If you change the negative hold time, the setup time is adjusted by
the intrument.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<HoldTime> Range: -99.999E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
TRIGger<m>:DATatoclock:STIMe
<SetupTime>
Sets the minimum time before the clock edge while the data signal must stay steady
above or below the data level.
The setup time can be negative. In this case, the hold time is always positive. The
setup/hold interval starts after the clock edge (setup time) and ends after the clock
edge (hold time). If you change the negative setup time, the hold time is adjusted by
the instrument.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<SetupTime> Range: -99.999E-9 to 100E-9
Increment: 1E-9
*RST: 0
Default unit: s
Usage: Asynchronous command
18.9.11 State trigger
The state trigger combines the edge trigger settings with trigger qualification. It is only
available for the A-event (Suffix = 1).
Use the following commands:
TRIGger<m>:EDGE:SLOPe on page 1168
TRIGger<m>:LEVel<n>[:VALue] on page 1166
TRIGger<m>:SCOupling on page 1184
TRIGger<m>:QUALify<n>:A[:ENABle] <State>
TRIGger<m>:QUALify<n>:B[:ENABle] <State>
Trigger

Table of Contents

Other manuals for Rohde & Schwarz RTE

Related product manuals