Remote control commands
R&S
®
RTE
1161User Manual 1326.1032.02 ─ 20
TRIGger<m>:RFSReject <HFRejectTrigger>
Enables the DSP filter for the trigger channel. Frequencies higher than the bandwidth
(TRIGger<m>:RFReject) are rejected, lower frequencies pass the filter.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<HFRejectTrigger> ON | OFF
*RST: OFF
Usage: Asynchronous command
18.8.11 Skew
CHANnel<m>:SKEW:MANual........................................................................................1161
CHANnel<m>:SKEW:TIME............................................................................................1161
PROBe<m>:SKEState.................................................................................................. 1162
CHANnel<m>:SKEW:MANual <ManualCompens>
If enabled, the skew offset value (CHANnel<m>:SKEW:TIME) is used for compensa-
tion. This improves horizontal and trigger accuracy.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ManualCompens> ON | OFF
*RST: ON
Usage: Asynchronous command
CHANnel<m>:SKEW:TIME <Offset>
Sets an delay value, that is known from the circuit specifics but cannot be compensa-
ted by the instrument automatically. It affects only the selected input channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Offset> Range: -100E-9 to 100E-9
Increment: 10E-15
*RST: 0
Default unit: s
Usage: Asynchronous command
Acquisition and setup