EasyManua.ls Logo

Rohde & Schwarz RTE - Skew

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1161User Manual 1326.1032.02 ─ 20
TRIGger<m>:RFSReject <HFRejectTrigger>
Enables the DSP filter for the trigger channel. Frequencies higher than the bandwidth
(TRIGger<m>:RFReject) are rejected, lower frequencies pass the filter.
Suffix:
<m>
.
1..3
The suffix is irrelevant.
Parameters:
<HFRejectTrigger> ON | OFF
*RST: OFF
Usage: Asynchronous command
18.8.11 Skew
CHANnel<m>:SKEW:MANual........................................................................................1161
CHANnel<m>:SKEW:TIME............................................................................................1161
PROBe<m>:SKEState.................................................................................................. 1162
CHANnel<m>:SKEW:MANual <ManualCompens>
If enabled, the skew offset value (CHANnel<m>:SKEW:TIME) is used for compensa-
tion. This improves horizontal and trigger accuracy.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<ManualCompens> ON | OFF
*RST: ON
Usage: Asynchronous command
CHANnel<m>:SKEW:TIME <Offset>
Sets an delay value, that is known from the circuit specifics but cannot be compensa-
ted by the instrument automatically. It affects only the selected input channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Offset> Range: -100E-9 to 100E-9
Increment: 10E-15
*RST: 0
Default unit: s
Usage: Asynchronous command