Remote control commands
R&S
®
RTE
1910User Manual 1326.1032.02 ─ 20
BUS<m>:PARallel:DECTable<n>:SHOW <ShowDecodeTable>
Available only if BUS<m>:PARallel:CLON is set to ON.
If enabled, a results box opens with decoded values of the bus signal and its time. You
can query the number of values (BUS<m>:PARallel:DECTable<n>:COUNt?) and
their data value (BUS<m>:PARallel:DECTable<n>:DATA?).
Suffix:
<m>
.
1..4
<n> 1..5
Parameters:
<ShowDecodeTable> ON | OFF
*RST: OFF
BUS<m>:PARallel:DECTable<n>:COUNt?
Returns the number of clock edges in the decode table.
Suffix:
<m>
.
1..4
<n> 1..5
Usage: Query only
BUS<m>:PARallel:DECTable<n>:DATA?
Returns the data of the decoded values of the bus signal and its time.
Suffix:
<m>
.
1..4
<n> 1..5
Usage: Query only
18.18.3 Digital resolution
ACQuire:DRESolution?.................................................................................................1910
ACQuire:POINts:DVALue?............................................................................................ 1911
ACQuire:DRESolution?
Returns the current digital resolution of the digital channels.
Return values:
<DigRes> Default unit: s
Usage: Query only
Asynchronous command
Mixed signal option (MSO, R&S RTE-B1)