Remote control commands
R&S
®
RTE
1187User Manual 1326.1032.02 ─ 20
DIRect
Input value remains unchanged
NOT
Input value is inverted
*RST: DIRect
Usage: Asynchronous command
TRIGger<m>:QUALify<n>:AB:LOGic <Operator>
TRIGger<m>:QUALify<n>:CD:LOGic <Operator>
TRIGger<m>:QUALify<n>:ABCD:LOGic <Operator>
Defines the logical combination of the indicated channels after evaluating the previous
logical operations:
AB: CH1 and CH3
CD: CH2 and CH4
ABCD: result of AB and CD
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
<n> 1..18
Trigger type: 10 = State, 11 = Pattern, all other suffixes are not
available
Parameters:
<Operator> AND | NAND | OR | NOR
AND
logical AND, conjunctive combination
NAND
logical NOT AND
OR
logical OR, disjunctive combination
NOR
logical NOT OR
*RST: AND
Usage: Asynchronous command
18.9.12 Pattern trigger
The pattern trigger is only available for the A-event (Suffix = 1).
The pattern is defined using the commands:
TRIGger<m>:QUALify<n>:A[:ENABle] on page 1185
TRIGger<m>:QUALify<n>:A:LOGic on page 1186
TRIGger<m>:QUALify<n>:AB:LOGic on page 1187