Search functions
R&S
®
RTE
453User Manual 1326.1032.02 ─ 20
Polarity, Range, Runt width, ±Delta
Time limit for the runt, see trigger settings:
●
"Polarity" on page 245
●
"Range" on page 248
●
"Runt width" on page 248
●
"±Delta" on page 248
Remote command:
SEARch:TRIGger:RUNT:POLarity on page 1359
SEARch:TRIGger:RUNT:RANGe on page 1360
SEARch:TRIGger:RUNT:WIDTh on page 1360
SEARch:TRIGger:RUNT:DELTa on page 1359
Upper level, Lower level
Set the upper and lower voltage thresholds. The amplitude of a runt crosses the first
threshold twice in succession without crossing the second one.
Remote command:
SEARch:TRIGger:LEVel:RUNT:UPPer on page 1361
SEARch:TRIGger:LEVel:RUNT:LOWer on page 1361
11.2.2.5 Window
The window search settings are the same as the window trigger settings. This search
type is not available if the search source is a digital channel. To apply the trigger set-
tings to search, tap Copy 'A Trigger'.
The window search is not available if the search source is a digital channel.
Search setup