Mixed signal option (MSO, R&S RTE-B1)
R&S
®
RTE
919User Manual 1326.1032.02 ─ 20
For the ranges "Within" and "Outside", the width defines the center of a range which is
defined by the limits "±Delta".
Remote command:
TRIGger<m>:PARallel:PATTern:WIDTh[:WIDTh] on page 1919
±Delta ← Width settings
Defines a range around the given width value.
The combination "Range" = Within and "±Delta" = 0 triggers on pulses with a pulse
width that equals "Width".
The combination "Range" = Outside and "±Delta" = 0 means to trigger on pulse widths
≠ "Width".
Remote command:
TRIGger<m>:PARallel:PATTern:WIDTh:DELTa on page 1920
14.3.1.8 Serial pattern
The serial pattern trigger identifies a serial bit string trigger on a single digital channel,
or for a logical combination of digital channels. The trigger requires a clocked bus; the
bits are read at the specified clock edge. The trigger event occurs at the last clock
edge of the serial bit string.
This trigger type allows you to trigger on specific address or data transmissions in
serial input and output signals.
Figure 14-7: Serial pattern trigger settings for trigger source = logical combination of digital channels
Clock edge
Sets the edge of the clock signal. The bit value is determined at the crossing of the
clock edge and the logical threshold.
Remote command:
TRIGger<m>:PARallel:SPATtern:CSOurce:EDGE on page 1920
Clock source
Selects the digital channel of the clock signal.
Remote command:
TRIGger<m>:PARallel:DATatoclock:CSOurce[:VALue] on page 1913
TRIGger<m>:PARallel:STATe:CSOurce:VALue on page 1913
TRIGger<m>:PARallel:SPATtern:CSOurce[:VALue] on page 1913
Trigger