Remote control commands
R&S
®
RTE
1113User Manual 1326.1032.02 ─ 20
Parameters:
<Mode> AUTO | OFF
AUTO: the instrument activates the roll mode under specific con-
ditions.
See also "Roll mode" on page 180.
*RST: AUTO
Usage: Asynchronous command
TIMebase:ROLL:STATe?
Returns the status of the roll mode.
Return values:
<State> ON | OFF
*RST: OFF
Usage: Query only
Asynchronous command
TIMebase:ROLL:MTIMe <MinHorizGain>
The roll mode is enabled automatically if the acquisition time exceeds the given value,
and if TIMebase:ROLL:ENABle is set to AUTO.
Parameters:
<MinHorizGain> Treshold value for roll mode enabling.
Range: 1 to 600
Increment: 1
*RST: 10
Default unit: s
Usage: Asynchronous command
18.8.3 Acquisition
CHANnel<m>[:WAVeform<n>][:STATe]........................................................................... 1113
CHANnel<m>[:WAVeform<n>]:TYPE..............................................................................1114
CHANnel<m>[:WAVeform<n>]:ARIThmetics....................................................................1114
ACQuire:COUNt...........................................................................................................1115
ACQuire:ARESet:IMMediate..........................................................................................1115
CHANnel<m>[:WAVeform<n>][:STATe] <State>
Activates or deactivates a waveform. [:STATe] can be omitted.
Suffix:
<m>
.
1..4
Selects the input channel.
Acquisition and setup