Remote control commands
R&S
®
RTE
1181User Manual 1326.1032.02 ─ 20
18.9.9 Slew rate trigger
TRIGger<m>:SLEW:SLOPe.......................................................................................... 1181
TRIGger<m>:LEVel<n>:SLEW:LOWer............................................................................1181
TRIGger<m>:LEVel<n>:SLEW:UPPer............................................................................ 1181
TRIGger<m>:SLEW:RANGe..........................................................................................1182
TRIGger<m>:SLEW:RATE............................................................................................ 1182
TRIGger<m>:SLEW:DELTa........................................................................................... 1183
TRIGger<m>:SLEW:SLOPe <Slope>
Selects the edge type for the trigger event.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
Parameters:
<Slope> POSitive | NEGative | EITHer
See Chapter 18.4.3, "Slope parameter", on page 1075.
*RST: POSitive
Usage: Asynchronous command
TRIGger<m>:LEVel<n>:SLEW:LOWer <Level>
TRIGger<m>:LEVel<n>:SLEW:UPPer <Level>
Set the lower and upper voltage thresholds. When the signal crosses a level, the slew
rate measurement starts or stops depending on the selected slope.
Suffix:
<m>
.
1..3
1 = A-trigger, 2 = B-trigger, 3 = R-trigger
<n> 1..31
Indicates the trigger source:
1...4 = channel 1...4
5...31 = not available
Parameters:
<Level> Range: -10 to 10
Increment: 1E-3
*RST: Lower = -0.1, upper =0.1
Default unit: V
Usage: Asynchronous command
Trigger