Getting started
R&S
®
RTE
68User Manual 1326.1032.02 ─ 20
"Reference point" defines the position of the reference point on the screen. Turn clock-
wise to move it to the right. The reference point marks the rescaling center of the time
scale. It is indicated by a gray triangle outline at the top of the diagram. If you modify
the time scale, the reference point remains fixed on the screen, and the scale is
stretched or compressed to both sides of the reference point.
[SCALE]
The rotary knob adjusts the time scale for all signals. The time scale is also known as
timebase.
Turn clockwise to stretch the waveforms. Doing so, the scale value time/div decreases.
Press the knob to toggle between coarse and fine scale adjustment.
3.2.3.4 VERTICAL controls
The keys and knobs in the VERTICAL functional block select a signal and adjust the
vertical scale and position of the selected signal.
[CH ×]
Turns on, selects, and configures a channel. If the channel is active, the key lights up
in the corresponding channel color .
The effect of the keypress depends on state of the channel:
●
If channel is off: Pressing the key turns on the channel and selects it.
●
If the channel is on, but not selected: Pressing the key selects the channel wave-
form.
●
If the waveform is selected: Pressing the key opens the "Vertical" dialog box for the
appropriate channel.
The vertical rotary knobs are focused on the selected waveform. They are illuminated
in the color of the selected waveform.
[REF]
Opens the "Reference" dialog box, where you can configure and display reference
waveforms. Press the key repeatedly to switch the reference waveform.
If a reference waveform is selected, the vertical rotary knobs are illuminated in white or
light gray (default colors), depending on the selected waveform.
Instrument tour