Mixed signal option (MSO, R&S RTE-B1)
R&S
®
RTE
921User Manual 1326.1032.02 ─ 20
14.4 Measurements on digital channels
For measurements on digital channels, the number of measurement categories and
types is reduced to applicable measurement.
These are:
Period
Frequency
Positive and negative pulse
Pulse count
Delay
Phase
Positive and negative duty cycle
Burst width
Edge count
Except for delay measurement, the measurements have the same settings and results
for analog and digital sources.
Delay measurement on digital channels is reduced to measure the time between two
subsequent rising or two subsequent falling edges.
See also Chapter 8.2, "Automatic measurements", on page 333.
14.5 Data export
The data of digital channels and parallel buses can be saved to file in the same way as
analog waveform data. One digital channel or bus per file can be saved.
The data format of the stored values is defined with "Analysis" > "Parallel Bus" >
"Setup" tab >"Data format".
The data format of the stored values is defined with "Analysis" > "Setup" > "Bus" tab
>"Data format".
Data export