Power analysis (option R&S RTE-K31)
R&S
®
RTE
955User Manual 1326.1032.02 ─ 20
16.1.1 General settings
Access: "Analysis"> "Power" > "General Settings"
The General settings, can be used by all measurements, like deskewing.
16.1.1.1 Auto deskew
The "Auto Deskew" dialog box guides you through the auto deskew of your current and
voltage probes.
Required equipment:
●
R&S RT-ZF20 power deskew fixture
●
Rohde & Schwarz voltage probe
●
Rohde & Schwarz current probe
1. Select "Analysis" > "Power".
2. Under "General", select "Auto Deskew".
3. Connect the voltage probe and the current probe to the oscilloscope.
4. Connect the probes to the R&S RT-ZF20 power deskew fixture as shown in the
"Channels" tab:
5. Select the correct channels for the "Current Source" and the "Voltage Source".
6. Tap "Next".
7. Check and complete the probe setup in the "Probes" tab.
Current probes and high-voltage differential probes are not automatically detected
by the instrument. Tap "Predefined probe" and select the correct probe type.
8. Tap the "Settings" tab.
Power measurement selection