Power analysis (option R&S RTE-K31)
R&S
®
RTE
957User Manual 1326.1032.02 ─ 20
Settings
In this tab, you can define the preset behavior and how the auto deskew results are
stored.
Current channel skew
Shows the skew settings of the channel connected to the current probe. Skew settings
are defined in the "Horizontal > Skew" dialog box.
See also: Chapter 5.9.1, "Skew", on page 232
Overwrite present skew setup
If disabled, the instrument only stores the result of the auto deskew procedure as a
separate value and does not use it. This value can be used later for power measure-
ments. The general skew offset under "Current channel skew" remains unchanged.
If enabled, the result of the auto deskew procedure is used for all measurements on
the selected channel. It is shown under "Current channel skew".
Remote command: 
POWer:DESKew:RESet on page 1945
Auto deskew result
Available only if "Overwrite present skew setup" is disabled.
"Deskew
value"
Result of the auto deskew.
Remote command: 
POWer:DESKew:TIME? on page 1945
"Apply"
Writes the result of the auto deskew to the "Skew offset" of the
selected channel.
Remote command: 
POWer:DESKew:CURRent on page 1944