Acquisition and waveform setup
R&S
®
RTE
181User Manual 1326.1032.02 ─ 20
Minimum acquisition time ← Roll mode
The instrument can activate the roll mode automatically if the Acquisition time exceeds
the value given here.
Remote command:
TIMebase:ROLL:MTIMe on page 1113
5.2.2 Acquisition
Access: [ACQUISITION] key
Acquisition settings control how the waveform is built from the captured samples.
For background information, see Chapter 5.1.2, "Sampling and acquisition",
on page 168.
The "Decimation" and "Arithmetic" settings are the same for all active channels.
Enable.........................................................................................................................181
Decimation.................................................................................................................. 181
Arithmetic.................................................................................................................... 182
Average count (N-single count)...................................................................................182
Reset...........................................................................................................................183
Enable
Activates or deactivates the channel.
Remote command:
CHANnel<m>[:WAVeform<n>][:STATe] on page 1113
Decimation
Selects the decimation mode. Decimation reduces the data stream of the ADC to a
stream of waveform points with lower sample rate and a less precise time resolution.
The R&S RTE uses decimation, if the waveform "Sample rate" is less than the ADC
sample rate. In this case, interpolation is not possible.
Horizontal settings