Waveform generator (Option R&S RTE-B6)
R&S
®
RTE
939User Manual 1326.1032.02 ─ 20
Content and format of the R&S arbitrary generator *.bin files
The file stream should contain the following information in the given order:
Sample rate
[double format]
Number of samples
[unsigned integer format]
Samples
[double format] * number of samples
15.1.5.1 General settings
Arb wfm source
Selects the arbitrary waveform source. You can load an existing file or load the current
oscilloscope waveform.
Remote command:
WGENerator<m>:ARBGen[:SOURce] on page 1935
Running mode
Selects the duration for which the signal of the arbitrary generator will be output after
the trigger event. You can choose between a "Continuous" and "Single period" dura-
tion.
Remote command:
WGENerator<m>:ARBGen:RUNMode on page 1934
Sample rate
Sets the sample rate for the arbitrary waveform.
Setup of the waveform generator