Triggers
R&S
®
RTE
243User Manual 1326.1032.02 ─ 20
signals related to the power line frequency, such as lighting equipment and power
supply devices.
●
Serial bus, D0...D15, Logic, Parallel bus 1...4:
If options with trigger functionality are installed, the variety of trigger sources is
enhanced with specific trigger sources.
Available sources depend on the trigger sequence setting. If "A only" is selected, all
inputs (analog input channels, serial and parallel buses, digital channels) can be used
as trigger source. If any other trigger sequence is selected, only channel inputs Ch1...4
can be set as trigger source, and all other input sources are disabled.
See also: Chapter 6.7, "Sequence", on page 270
Remote command:
TRIGger<m>:SOURce[:SELect] on page 1165
Type
Selects the trigger type specific for each condition in a trigger sequence. The current
trigger type is shown on the button.
The following trigger types are available:
●
Edge, see page 243
●
Glitch, see page 244
●
Width, see page 245
●
Runt, see page 247
●
Window, see page 248
●
Timeout, see page 250
●
Interval, see page 251
●
Slew rate, see page 252
●
Data2Clock trigger, see page 254
●
State, see page 255
●
Pattern, see page 256
●
Serial pattern, see page 258
●
TV/Video trigger, see page 259
●
Line trigger, see page 263
Restrictions:
●
If the external trigger input is used as trigger source, the analog edge trigger is the
only available trigger type.
Remote command:
TRIGger<m>:TYPE on page 1166
Find level
Sets the trigger level automatically to 0.5 * (MaxPeak – MinPeak). The function is not
available for an external trigger source and the TV trigger.
Remote command:
TRIGger<m>:FINDlevel on page 1167
6.3.2 Edge
Access: [TRIGGER] > "Setup" tab > "Type = Edge"
Trigger types