Remote control commands
R&S
®
RTE
1844User Manual 1326.1032.02 ─ 20
Suffix:
<m>
.
1..4
<n> *
Return values:
<Stop> Range: -100E+24 to 100E+24
Increment: 100E-12
*RST: 0
Default unit: s
Usage: Query only
BUS<m>:USBPd:RESult:FRAMe<n>:TYPE?
Returns the frame type for the selected serial bus and frame number.
Suffix:
<m>
.
1..4
<n> *
Return values:
<FrameType> STARt | RESet | CTRL | DATA | BIST | RQST | SINK | SRC |
VEND | BATT | ALRT | XMSG | XSRC | XSTA | XGBC | XGBS |
XBAC | XGMI | XMFI | XSRQ | XSRS | XFRQ | XFRS | TEST |
LOWP | UNKNown
For a description of the frame types, see "Frame packet types"
on page 849.
*RST: UNKNown
Usage: Query only
18.17.18.4 Search settings
The trigger suffix <m> is always 1 and can be omitted. It selects the trigger event: Only
the A-trigger is available for triggering on serial buses.
To trigger on a serial bus, make sure that:
●
TRIGger<m>:SOURce[:SELect] is set to SBUS.
●
The sources of the serial bus are channel signals: use BUS<m>:...:SOURce
commands.
●
Decoding is enabled: BUS<m>[:STATe] is set to ON.
There are two commands for each parameter, that you can use for defining the USBPD
settings.
For example, to set the Frame type =Control Frame > Field =MsgID >Data value you
can use one of the following commands:
●
SEARch:TRIGger:USBPD:FRAMe3:FLD3:DMIN 01100
Defines the parameter by using the index <m> for the frame number and <n> for
the field number. For an overview, see Table 18-12.
Protocols