EasyManua.ls Logo

Rohde & Schwarz RTE - Noise Rejection

Rohde & Schwarz RTE
2078 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commands
R&S
®
RTE
1382User Manual 1326.1032.02 ─ 20
Example: Prerequisite: Search1 and Zoom2 are available.
SEARch:GATE:ZCOupling 'Search1',ON
SEARch:GATE:ZDIagram 'Search1','Diagram1','Zoom2'
SEARch:GATE:ZDIagram? 'Search1'
<-- Diagram1;Zoom2
SEARch:GATE:STATe 'Search1', ON
SEARch:GATE:SHOW 'Search2', ON
Enables the zoom coupling to define the gate,and selects
Zoom2 as gate area. Zoom2 is based on Diagram1. Then the
gate is activated and displayed.
Usage: Asynchronous command
18.15.16
Noise rejection
SEARch:TRIGger:LEVel:NOISe:ABSolute...................................................................... 1382
SEARch:TRIGger:LEVel:NOISe:MODE.......................................................................... 1382
SEARch:TRIGger:LEVel:NOISe:RELative.......................................................................1383
SEARch:TRIGger:LEVel:NOISe[:STATe].........................................................................1383
SEARch:TRIGger:LEVel:NOISe:ABSolute <SearchName>, <SignalSource>,
<Value>
Defines the trigger hysteresis, a range in absolute values around the trigger level. If the
signal jitters inside this range and crosses the trigger level, no trigger event is detected.
Parameters:
<Value> Hysteresis value
Setting parameters:
<SearchName> String with the name of the search
<SignalSource> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | TRK1 | TRK2 |
TRK3 | TRK4 | TRK5 | TRK6 | TRK7 | TRK8 | Z1V1 | Z1V2 |
Z1V3 | Z1V4 | Z1I1 | Z1I2 | Z1I3 | Z1I4 | Z2V1 | Z2V2 | Z2V3 |
Z2V4 | Z2I1 | Z2I2 | Z2I3 | Z2I4
Source of the search, see Chapter 18.4.2, "Waveform parame-
ter", on page 1074
Usage: Asynchronous command
SEARch:TRIGger:LEVel:NOISe:MODE <SearchName>, <SignalSource>, <Mode>
SEARch:TRIGger:LEVel:NOISe:MODE? <Key>, <SignalSource>
Defines whether absolute values or relative values to the vertical scaling are used as a
hysteresis for noise rejection.
Parameters:
<Mode> ABS | REL
Search

Table of Contents

Other manuals for Rohde & Schwarz RTE

Related product manuals