Remote control commands
R&S
®
RTE
1985User Manual 1326.1032.02 ─ 20
18.22 Status reporting
This chapter describes the remote commands that are used to read the status regis-
ters.
For information on structure, hierarchy, and contents of the status registers, see Chap-
ter 17.5.4, "Remote control - status reporting system", on page 1046.
● STATus:OPERation register................................................................................1985
● STATus:QUEStionable registers......................................................................... 1986
18.22.1 STATus:OPERation register
STATus:OPERation commands provide information on the activity of the instrument.
See also: "STATus:OPERation register" on page 1047
STATus:OPERation:CONDition?....................................................................................1985
STATus:OPERation[:EVENt]?........................................................................................1985
STATus:OPERation:ENABle..........................................................................................1985
STATus:OPERation:CONDition?
STATus:OPERation[:EVENt]?
The CONDition command returns information on actions the instrument is currently
executing. The contents of the register is retained.
The EVENt command returns information on actions the instrument has executed
since the last reading. Reading the EVENt register deletes its contents.
Bits:
●
0 = ALIGnment
●
2 = AUToset
●
3= WTRIgger (wait for trigger)
●
4= MEASuring
Usage:
Query only
STATus:OPERation:ENABle <Enable>
Controls the ENABle part of the STATus:OPERation register. The ENABle defines
which events in the EVENt part of the status register are forwarded to the OPERation
summary bit (bit 7) of the status byte. The status byte can be used to create a service
request.
Parameters:
<Enable> Range: 1 to 65535
Increment: 1
Example:
STATus:OPERation:ENABle 5
The ALIGnment event (bit 0) and AUToset event (bit 2) are for-
warded to the OPERation summary bit of the status byte.
Status reporting