Waveform generator (Option R&S RTE-B6)
R&S
®
RTE
949User Manual 1326.1032.02 ─ 20
● Configuring a function waveform...........................................................................949
● Configuring a modulation waveform......................................................................949
● Configuring a sweep waveform.............................................................................951
● Configuring an arbitrary waveform........................................................................951
● Configuring a pattern generator waveform............................................................952
15.4.1  Configuring a function waveform
1. Press the [Gen 1] key on the front panel.
2. Select the "Setup" tab.
3. Under "Operation mode", enable the "Function Gen." button.
4. Select the "Function type" that you want to generate, e.g. "Sine".
5. Depending on the selected "Function type", configure the settings of the waveform
like "Frequency" and "Period".
6. If necessary, change the "Output" settings or add "Noise" to the waveform.
7. Press the "Enable" button, to output the waveform at the output connector of the
waveform generator.
15.4.2  Configuring a modulation waveform
Generating an AM modulated waveform
1. Press the [Gen 1] key on the front panel.
2. Select the "Setup" tab.
3. Under "Operation mode", enable the "Modulation" button.
4. Tap "Modulation type" and select "AM".
5. Set the "Carrier frequency" and the "Carrier period".
6. Tap "Modulation signal" and select the required waveform.
7. Depending on the selected "Modulation signal", configure the settings of the wave-
form like "Modulation freq" and "Modulation depth".
8. If necessary, change the "Output" settings or add "Noise" to the waveform.
9. Press the "Enable" button, to output the waveform at the output connector of the
waveform generator.
Generating an PWM modulated waveform
1. Press the [Gen 1] key on the front panel.
2. Select the "Setup" tab.
Configuring the waveform generator