Getting started
R&S
®
RTE
76User Manual 1326.1032.02 ─ 20
3.3.2 Acquiring data
You can acquire data using different arithmetic methods to get envelope or average
waveforms.
If you need to change only one setting during analysis, and you need to change it
often, you can reduce the dialog box to a small box that only contains the required set-
ting. Thus you can change the setting and see the result immediately.
1. Press the [ACQUISITION] key on the front panel, in the [HORIZONTAL] area.
2. In the "Acquisition" tab of the "Horizontal" dialog box, select the "Decimation" type
Sample.
3. Tap the "Arithmetic" button for C1 but do not change the setting.
4.
Tap the
"Minimize" icon in the upper right corner of the dialog box.
The dialog box turns into a mini box that contains only the "Wfm Arithmetic" setting.
5. Select Envelope and check the waveform.
The instrument sets the "Decimation" type automatically to Peak detect to display
the correct envelope waveform.
6. Select Average and check the waveform.
7.
Tap the
"Maximize" icon in the mini dialog box.
The complete "Acquisition" dialog box is restored.
8.
Close the dialog box by tapping .