Remote control commands
R&S
®
RTE
1311User Manual 1326.1032.02 ─ 20
Usage: Asynchronous command
18.12.13.3 Manual configuration
REFLevel<m>:ABSolute:HIGH...................................................................................... 1311
REFLevel<m>:ABSolute:LOW....................................................................................... 1311
REFLevel<m>:ABSolute:TDIStance............................................................................... 1311
REFLevel<m>:ABSolute:BDIStance...............................................................................1312
REFLevel<m>:ABSolute:MLEVel................................................................................... 1312
REFLevel<m>:RELative:UPPer..................................................................................... 1313
REFLevel<m>:RELative:MIDDle....................................................................................1313
REFLevel<m>:RELative:LOWer.................................................................................... 1313
REFLevel<m>:ABSolute:HIGH <Value>
The signal value that represents a high level.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 111, 115 to 118
Source waveform of the measurement, see Chapter 18.4.1,
"Waveform suffix", on page 1072.
Parameters:
<Value>
Example:
REFLevel2:ABSolute:HIGH 0.015
Sets the high signal level for Ch1 to 15 mV. C1W1 corresponds
to suffix number 2.
Usage: Asynchronous command
REFLevel<m>:ABSolute:LOW <Value>
The signal value that represents a low level.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 111, 115 to 118
Source waveform of the measurement, see Chapter 18.4.1,
"Waveform suffix", on page 1072.
Parameters:
<Value>
Example:
REFLevel2:ABSolute:Low 0.0015
Sets the low signal level forCh1 to 1.5 mV. C1W1 corresponds
to suffix number 2.
Usage: Asynchronous command
REFLevel<m>:ABSolute:TDIStance <Value>
The distance between the high signal level and the upper reference level.