Remote control commands
R&S
®
RTE
1313User Manual 1326.1032.02 ─ 20
Example: See: "Manual reference level definition using absolute values"
on page 1058
Usage: Asynchronous command
REFLevel<m>:RELative:UPPer <Value>
Sets the upper relative reference level if REFLevel<m>:RELative:MODE is set to
USER.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 111, 115 to 118
Source waveform of the measurement, see Chapter 18.4.1,
"Waveform suffix", on page 1072.
Parameters:
<Value> Percentage of the high signal level.
Example:
REFLevel8:RELative:LOWer 85
Sets the upper reference level for Ch3 to 85 %. Ch3 corre-
sponds to suffix number 8.
Example: See: "Manual reference level definition using relative values"
on page 1057
Usage: Asynchronous command
REFLevel<m>:RELative:MIDDle <Value>
Sets the middle relative reference level if REFLevel<m>:RELative:MODE is set to
USER.
Suffix:
<m>
.
Valid suffix numbers: 2 to 21, 61 to 68, 108 to 111, 115 to 118
Source waveform of the measurement, see Chapter 18.4.1,
"Waveform suffix", on page 1072.
Parameters:
<Value> Percentage of the high signal level.
Example:
REFLevel8:RELative:MIDDle 50
Sets the middle reference level for Ch3 to 50 %. Ch3 corre-
sponds to suffix number 8.
Example: See: "Manual reference level definition using relative values"
on page 1057
Usage: Asynchronous command
REFLevel<m>:RELative:LOWer <Value>
Sets the lower relative reference level if REFLevel<m>:RELative:MODE is set to
USER.
Automatic measurements