Triggers
R&S
®
RTE
250User Manual 1326.1032.02 ─ 20
"Longer"
Triggers if the signal crosses vertical limits after the specified "Width"
time is reached.
Remote command:
TRIGger<m>:WINDow:TIME on page 1177
Width
For the ranges "Within" and "Outside", the width defines the center of a time range
which is defined by the limits "±Delta".
For the ranges "Shorter" and "Longer", it defines the maximum and minimum time
lapse, respectively.
Remote command:
TRIGger<m>:WINDow:WIDTh on page 1177
±Delta
Defines a range around the "Width" value.
Remote command:
TRIGger<m>:WINDow:DELTa on page 1178
6.3.7 Timeout
Access: [TRIGGER] > "Setup" tab > "Type = Timeout"
The timeout trigger checks if the signal stays above or below the threshold voltage for
a specified time lapse. In other words, the trigger occurs if the trigger source does not
have the expected transition within the specified time.
Trigger level
Sets the voltage level for the trigger condition. You can also drag the trigger level
marker on the display (TA or TB on the right edge of the display). The range of the trig-
ger level is limited in a way so that always a hysteresis for stable trigger conditions is
available.
Remote command:
TRIGger<m>:LEVel<n>[:VALue] on page 1166
Range
Selects the relation of the signal level to the trigger level: