Protocol analysis
R&S
®
RTE
851User Manual 1326.1032.02 ─ 20
Hysteresis
Sets a value for the hysteresis of the data.
Remote command:
BUS<m>:USBPd:HYSTeresis on page 1832
Current advertisment thresholds
The signal level provides information about the current advertisement between the
bursts. These thresholds determine the levels at which the current advertisement
modes are defined.
"vRd-1.5"
Sets the threshold at USB Type-C current of 1.5 A.
"vRd-USB"
Sets the threshold at default USB Type-C current.
"vRa"
Sets the threshold for the low current.
Remote command:
BUS<m>:USBPd:THRBottom on page 1832
BUS<m>:USBPd:THRMid on page 1833
BUS<m>:USBPd:THRTop on page 1833
Show decode detail
If enabled, the data words are broken down into subframes. If not enabled the data
words are displayed as 32-bit data words.
If the "Show decode detail" is enabled, you cannot do a search and trigger on USBPD
frames.
Remote command:
BUS<m>:USBPd:DETail on page 1831
13.17.2.2 Display settings
Access: [PROTOCOL] > "Display" tab.
For common display settings, see Chapter 13.1.3, "Display", on page 515.
To enhance the decode possibilities of the USBPD protocol, you can use an additional
setting in the "Display" tab: "Decode layer".
USBPD (option R&S RTE-K63)