Measurements
R&S
®
RTE
341User Manual 1326.1032.02 ─ 20
Default position
Defines the default position of the result table, when a new measurement is started:
"Docked": fixed tab below the diagrams
"Preview": result icon on the signal bar
"Floating": floating result box in front of the diagrams
Remote command:
DISPlay:RESultboxes:MEPosition on page 1288
Show statistic values
Select the statistical values that you want to see in the results table.
8.2.3 Measurement gates
Gate areas limit the measurement to a user-defined range of the waveform. The gate
settings are defined on the "Gate " tab.
Each measurement can use its own gate. Make sure to select the correct measure-
ment.
8.2.3.1 Using measurement gates
If you have enabled the measurement with the toolbar icon and drew a rectangle on
the diagram, the gate is automatically defined and enabled. If you want to create a gate
for a running measurement, or if you want to modify the gate area, configuration is
done in the "Measurements" > "Gate" dialog box.
1. On the "Meas" menu, tap "Gate/Display".
2. Select the subtab for the measurement you want to configure.
3. To define the gate, use one of the following methods:
Define the start value and the stop value of the gate area by entering either
absolute or relative values.
If a zoom area has already been defined for the waveform, couple the gate
area to the zoom area by selecting the "Zoom coupling" option.
If a cursor measurement has already been defined for the waveform, couple
the gate area to the cursor lines by selecting the "Cursor coupling" option.
4. If you want to use the same gate for all measurement, enable "Use gate coupling".
5. Tap the "Use gate" icon to enable the gate usage.
The measurement is performed on the selected part of the waveform. The gate is
shown in the diagram.
8.2.3.2 Gate settings for measurements
Access: "Meas" menu > "Gate/Display" tab
Automatic measurements